INVESTIGADORES
PACHECO Pablo Hugo
artículos
Título:
Development of an on line miniaturized non-chromatographic arsenic speciation system
Autor/es:
ARIEL MARATTA; LUIS DANTE MARTINEZ; PABLO PACHECO
Revista:
MICROCHEMICAL JOURNAL
Editorial:
ELSEVIER SCIENCE BV
Referencias:
Lugar: Amsterdam; Año: 2016 vol. 127 p. 199 - 205
ISSN:
0026-265X
Resumen:
An arsenic speciation method is proposed introducing selective hydride generation, UV photooxidation and hydride trapping on oxidized multiwall carbon nanotubes (CNTs) with electrothermal atomic absorption spectrometry (ETAAS) for detection. To this end a flow injection system was designed to focus on miniaturization with minimal reagent consumption. As a result reagent consumption was kept in μL level with no need of carrier gas for hydride transport. This study obtained an enhancement factor of 60 when 2 mL of sample were loaded in the gas?liquid separator, arsine adsorbed on CNTs, and eluted with 30 μL of 5% HNO3 (v v− 1). A limit of detection (LoD) of 0.78 ng L− 1 was obtained. The precision was evaluated by relative standard deviation (RSD%) corresponding to 8.3% (n = 10). The method determined organic and inorganic arsenic fractions; and As3 + and As5 + concentrations in well and cistern water samples from arsenic endemic regions of Argentina.