Título:
On-line solid phase extraction CZE for the simultaneous determination of lanthanum and gadolinium at picogram per liter levels
Autor/es:
NORA VIZIOLI; RAÚL A. GIL; LUIS D. MARTINEZ; MARIA F. SILVA
Revista:
ELECTROPHORESIS
Editorial:
WILEY-V C H VERLAG GMBH
Referencias:
Año: 2009 p. 2681 - 2687
ISSN:
0173-0835
Resumen:
A non-specific on-line method is presented for the extraction and preconcentration of two rare earth elements using a microcartridge containing C18-derivatized silica particles prior to their analysis by CZE. The microcartridge, named analyte concentrator, was coupled on-line to the inlet of the separation capillary (fused-silica (FS) capillary, 75 µm id ×12 cm from the inlet to the microcartidge and 37 cm from the microcartridge to the detector). The reversed-phase sorbent quantitatively retained gadolinium (Gd) and lanthanum (La) as 2-(5-bromo-2-pyridylazo)-5-diethylaminophenol complexes in the presence of non-ionic micelles of polyethylene glycol tert-octylphenyl ether, enabling sample clean-up and concentration enhancement with minimum sample handling. The rare earth elements chelates were released from the sorbent with methanol and then analyzed by CZE with diode array detection. A background electrolyte of 20 mM sodium tetraborate containing 8% ACN, pH 9.0, was found to be optimal for the separation of metal chelates. The concentration limits of detection were lowered to picogram per liter levels (20 pg/L for La and 80 pg/L for Gd). A 1000-fold improvement in concentration sensitivity for La- and Gd-2-(5-bromo-2-pyridylazo)-5-diethylaminophenol complexes with respect to CZE without preconcentration was reached.