Título:
Tetradecyl(trihexyl)phosphonium chloride ionic liquid single-drop microextraction for electrothermal atomic absorption spectrometric determination of lead in water samples
Autor/es:
MARTINIS, E.M.; BERTON, P.; ALTAMIRANO, J.C.; HAKALA, U; WUILLOUD, R.G.
Revista:
TALANTA
Editorial:
ELSEVIER SCIENCE BV
Referencias:
Año: 2009 vol. 80 p. 2034 - 2040
ISSN:
0039-9140
Resumen:
A highly efficient single drop microextraction (SDME) procedure using a low cost room temperature ionic liquid (RTIL), i.e. tetradecyl(trihexyl)phosphonium chloride (CYPHOSÒ IL 101), for Pb determination at trace levels in real water samples was developed. Lead was chelated with 2-(5-bromo-2-pyridylazo)-5-diethylaminophenol (5-Br-PADAP) reagent and extracted into a 4 μL micro-drop of CYPHOSÒ IL 101. The RTIL drop was directly injected into the graphite furnace of the electrothermal atomic absorption spectrometer (ETAAS). Under optimal microextraction conditions, a preconcentration factor of 32 was achieved with only 1.5 mL of sample resulting in a phase-volume ratio of 375. The limit of detection (LOD) obtained was 3.2 ng L-1 and the relative standard deviation (RSD) for 10 replicates at 0.5 mg L-1 Pb2+ concentration level was 4.9%, calculated at peak heights. The calibration graph was linear from concentration levels near the detection limits up to at least 4.5 mg L-1 with a correlation coefficient of 0.9996. The accuracy of the methodology was evaluated by analysis of a certified reference material (CRM). The method was successfully applied to the determination of Pb in tap, mineral, well and river water samples.