Design of a Cooper pair box electrometer for application to solid-state and astroparticle physics
Autor/es:
HELENE LE SUEUR; DENIS VION; LEANDRO TOSI
Revista:
PHYSICAL REVIEW APPLIED
Editorial:
AMERICAN PHYSICS SOCIETY
Referencias:
Año: 2019 vol. 11
ISSN:
2331-7019
Resumen:
We describe the design and principle of operation of a fast and sensitive electrometer operated at millikelvin temperatures, which aims at replacing conventional semiconducting charge amplifiers in experiments needing low back-action or high sensitivity. The electrometer consists of a Cooper Pair box (CPB) coupled to a microwave resonator, which converts charge variations to resonance frequency shifts. We analyze the dependence of the sensitivity on the various parameters of the device, and derive their optimization. By exploiting the nonlinearities of this electrometer, and using conventional nanofabrication and measurement techniques, a charge sensitivity of a few 10−7e/Hz−−−√ can be achieved which outperforms existing single charge electrometers.