Automatic digital-analog impedance plethysmograph
Autor/es:
YANICELLI, LUCÍA MARÍA; MAURO, KEVIN ALFONSO; HERRERA, MYRIAM CRISTINA; MARÍA DE LOS ANGELES, GÓMEZ LÓPEZ; GOY, CARLA BELÉN; PARODI, NICOLAS FRANCO
Revista:
Journal of Physics: Conference Series
Editorial:
IOP Publishing (UK)
Referencias:
Año: 2016
ISSN:
1742-6588
Resumen:
Venous occlusion plethysmography (VOP) is a traditional method widely used to assess limb blood circulation. One common mode to record VOP is by means of evaluating limb volume changes using impedance plethysmography (IP). In this paper the design and implementation of an automatic digital-analog impedance plethysmograph (ADAIP) for VOP is presented. The system is tested using precision resistances in order to calculate its repeatability. Then its global performance is assessed by means of VOP recordings on the upper and me lower limb of a healthy volunteer. The obtained repeatability was very high (95%), and the VOP recordings where the expected ones. It can be concluded that the whole system performs well and that it is suitable for automatic VOP recording.