INVESTIGADORES
COPELLO Guillermo Javier
artículos
Título:
Analysis of Diphenylamine and Impurities Using Monolithic Column with Electrochemical Detection
Autor/es:
S. L. IGLESIAS; M. F. DESIMONE; G. J. COPELLO; J. A. BERTINATTO; S. A. GIORGIERI; L. E. DÍAZ
Revista:
JOURNAL OF ANALYTICAL CHEMISTRY
Editorial:
Springer
Referencias:
Año: 2006 vol. 61 p. 588 - 591
ISSN:
1061-9348
Resumen:
A high-performance liquid chromatographic (HPLC) method with electrochemical detection (ED) using a monolithic column was developed and validated for the simultaneous quantitative determination of aniline, 2-aminobiphenyl, and 4-aminobiphenyl in diphenylamine samples. The proposed HPLC method provides a shorter analysis time as compared to previously reported ones (less than 4 min). Method validation was performed in terms of sensitivity, repeatability, linearity, accuracy, detection limit, and quantification limit. It is found that the method has high accuracy and precision. The limits of detection obtained were 0.23 ng/mL for aniline and 0.30 ng/mL for 2-aminobiphenyl, 4-aminobiphenyl, and diphenylamine.