Título:
A fully automated system for inorganic antimony preconcentration and speciation in urine
Autor/es:
PABLO H. PACHECO, RAÚL A. GIL, LUIS D. MARTINEZ, GRISELDA POLLA, PATRICIA SMICHOWSKI
Revista:
ANALYTICA CHIMICA ACTA
Editorial:
ELSEVIER SCIENCE BV
Referencias:
Año: 2007 p. 1 - 7
ISSN:
0003-2670
Resumen:
A studywas undertaken to ascertain the analytical capabilities of l-methionine immobilized
on controlled pore glass for Sb preconcentration and speciation. A fully automated on-line
system, implemented with hydride generation (HG) and inductively coupled plasma optical
emission spectrometry (ICP OES), was used. Sb(III), at pH 10 was selectively retained in
the column containing the immobilized aminoacid, while Sb(V) was not retained at all. A
30% HCl solution was used as eluent agent. Prior to total Sb determination, a pre-reduction
step with thiourea was necessary. An on-line pH adjusting and pre-reduction of Sb(V) was
achieved in a fully automated system. The detection limit for the preconcentration of 10mL
of an aqueous solution was 70ng L-1 with a relative standard deviation of 2%. An enrichment
factor of 20 was achieved when 10mL of sample was passed through the system,
reaching a throughput of 23 samples per hour. The method was successfully applied to the
determination of Sb(III) and total Sb in urine.l-methionine immobilized
on controlled pore glass for Sb preconcentration and speciation. A fully automated on-line
system, implemented with hydride generation (HG) and inductively coupled plasma optical
emission spectrometry (ICP OES), was used. Sb(III), at pH 10 was selectively retained in
the column containing the immobilized aminoacid, while Sb(V) was not retained at all. A
30% HCl solution was used as eluent agent. Prior to total Sb determination, a pre-reduction
step with thiourea was necessary. An on-line pH adjusting and pre-reduction of Sb(V) was
achieved in a fully automated system. The detection limit for the preconcentration of 10mL
of an aqueous solution was 70ng L-1 with a relative standard deviation of 2%. An enrichment
factor of 20 was achieved when 10mL of sample was passed through the system,
reaching a throughput of 23 samples per hour. The method was successfully applied to the
determination of Sb(III) and total Sb in urine.-1 with a relative standard deviation of 2%. An enrichment
factor of 20 was achieved when 10mL of sample was passed through the system,
reaching a throughput of 23 samples per hour. The method was successfully applied to the
determination of Sb(III) and total Sb in urine.