INVESTIGADORES
MOLINA Pablo Fernando
congresos y reuniones científicas
Título:
A tissue-equivalent radioluminescent fiberoptic probe for in-vivo dosimetry based on Mndoped lithium tetraborate
Autor/es:
M SANTIAGO; M PROKIC; P MOLINA; J MARCAZZÓ; E CASELLI
Lugar:
Munich
Reunión:
Congreso; ,11th World Congress on Medical Physics and Biomedical Engineering; 2009
Institución organizadora:
International Union for Physical and Engineering Sciences in Medicine (IUPESM)
Resumen:
The fiberoptic (FO) dosimetry concept, which relies on the assessment of dose by measuring the intensity of the light emitted by a tiny sample of a radioluminescent (RL) compound coupled to an optical fiber cable, is gaining importance as a promising dosimetry technique for in-vivo dosimetry. In this work the design, construction and test of a tissue-equivalent RL probe is described. The intensity of the RL emission of Mn-doped Lithium Tetraborate samples during irradiation is used as an estimate of the relative dose. The influence of spurious luminescence is discussed and a removing method based on simple optical filtering is implemented. Finally, the response of the probe in the context of a typical depth-dose experiment in a 60Co radiotherapy facility is analyzed and compared to the response of a standard ionizing chamber.