IAFE   05512
INSTITUTO DE ASTRONOMIA Y FISICA DEL ESPACIO
Unidad Ejecutora - UE
artículos
Título:
Multi-wavelength calibration procedure for the Pierre Auger Observatory fluorescence detectors
Autor/es:
ROVERO, A.C., BAULEO, P., BRACK, J., HARTON, J., KNAPIK, R., FOR THE PIERRE AUGER COLLABORATION
Revista:
ASTROPARTICLE PHYSICS
Editorial:
Elsevier
Referencias:
Lugar: Holanda; Año: 2009 vol. 31 p. 305 - 311
ISSN:
0927-6505
Resumen:
We present a method to measure the relative spectral response of the Pierre Auger Observatory Fluorescence Detector. The calibration was done at wavelengths of 320, 337, 355, 380 and 405 nm using an end-to-end technique in which the response of all detector components are combined in a single measurement. A xenon flasher and notch-filters were used as the light source for the calibration device. The overall uncertainty is 5%.