INSIBIO   05451
INSTITUTO SUPERIOR DE INVESTIGACIONES BIOLOGICAS
Unidad Ejecutora - UE
artículos
Título:
Design and Evaluation of Pressure and Eelectrocardiographic Modules Embedded in a Cardiac Hemodynamic Monitor
Autor/es:
CÓRDOBA, LUIS P.; GOMEZ LOPEZ, MARIA DE LOS ANGELES; GOY, CARLA BELÉN; HERRERA, MYRIAM CRISTINA
Revista:
IEEE LATIN AMERICA TRANSACTIONS
Editorial:
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
Referencias:
Lugar: New York; Año: 2013 vol. 11 p. 107 - 113
ISSN:
1548-0992
Resumen:
This work presents the design and the implementation of a two-channels recording system and its adaptation to a FPGA-based acquisition system. The recorded signals are Electrocardiogram (ECG) and Intraventricular Pressure (IVP). Elements and circuits needed to incorporate the newly developed recording system to a previously developed FPGA-based Conductance Catheter System (SCC) are also presented. The result is a fully integrated system, in which the evaluation of each channel is carried out. ECG amplifier is evaluated by means of frequency response and common mode rejection ratio. Filter frequency response and two calibration methods, over 0-150 mmHg and 150-300 mmHg ranges, are used to evaluate PIV channel. Finally analogical to digital conversion and communication between two-channels recording system and FPGA-based acquisition system are tested. Satisfactory results are obtained.