Título:
Experimental and Simulated Characterization of a Beam Shaping Assembly for Accelerator- Based Boron Neutron Capture Therapy (ABBNCT)
Autor/es:
A.A. BURLON; S. GIROLA; A.A. VALDA; D.M. MINSKY; A.J. KREINER
Lugar:
Santiago, Chile
Reunión:
Congreso; VIII Latin American Symposium on Nuclear Physics and Applications; 2009
Institución organizadora:
Facultad de Ciencias Físicas y Matemáticas, Universidad de Chile
Resumen:
Artículo completo enviado - en proceso de evaluación - n the frame of the construction of a Tandem Electrostatic Quadrupole Accelerator facility devoted to the Accelerator-Based Boron Neutron Capture Therapy, a Beam Shaping Assembly has been characterized by means of Monte-Carlo simulations and measurements. The neutrons were generated via the 7Li(p,n)7Be reaction by irradiating a thick LiF target with a 2.3 MeV proton beam delivered by the TANDAR accelerator at CNEA. The emerging neutron flux was measured by means of activation foils while the beam quality and directionality was evaluated by means of Monte Carlo simulations. The parameters show compliance with those suggested by IAEA. Finally, an improvement adding a beam collimator has been evaluated.