INVESTIGADORES
GARCIA Guillermo Manuel
artículos
Título:
Comparison of the MagNA Pure LC Automated System and the Ribo Pure – Blood RNA Manual Method for RNA Extraction from Multiple Myeloma Bone Marrow Samples Conserved in an RNA stabilizer.
Autor/es:
GARCÍA-EFFRON, GUILLERMO; GAMARRA, SOLEDAD; CROOKE, ALVAREZ ALMUDENA; MARTINEZ-SANCHEZ, PILAR; LAHUERTA-PALACIOS, JUAN JOSÉ; MARTÍNEZ-LÓPEZ, JOAQUÍN
Revista:
International journal of laboratory hematology
Editorial:
Oxford Journals
Referencias:
Lugar: Oxford (Reino Unido); Año: 2007 vol. 29 p. 139 - 144
ISSN:
1751-5521
Resumen:
A total of 62 frozen bone marrow specimens conserved in RNA later (Ambion) were processed using two different extraction methods, the MagNA Pure LC system (MAG; Roche) and the manual RiboPure-Blood RNA method (RIBO; Ambion); Beta glucoronidase RNA (GUS) was amplified by LightCycler PCR to evaluate the quality of both extraction procedures. Less than 1000 GUS copies/ml was detected in 26 of 62 specimens (41.94%) processed by MAG and in five of 62 specimens (8.06%) processed by RIBO. Moreover, RNA recovery from the 62 specimens by MAG is, on average, 2.91 cycle threshold-fold higher than RIBO (P = 0.0008). Furthermore, we compared the extraction times and reagent costs of both methods. In conclusion, RNA extraction using MAG is faster to process 32 samples and less expensive than RIBO but it is not sensitive enough to be employed for research purpose in our laboratory.