IITEMA   27339
INSTITUTO DE INVESTIGACIONES EN TECNOLOGIAS ENERGETICAS Y MATERIALES AVANZADOS
Unidad Ejecutora - UE
artículos
Título:
An inner filter effect based sensing system for the determination of caffeine in beverage samples
Autor/es:
ZANINI, GRACIELA; LANTMANN CORRAL, SARA P.; ACEBAL, CAROLINA C.; GONZÁLEZ, NATALIA; MONTEJANO, HERNÁN
Revista:
ANALYST
Editorial:
ROYAL SOC CHEMISTRY
Referencias:
Año: 2020
ISSN:
0003-2654
Resumen:
In this work, the inner filter effect (IFE) of caffeine (CF) over the fluorescence signal of glibenclamide (GLB) was used for the determination of CF in beverage samples. The system worked in a turn-off mode since the absorption spectrum of CF overlaps the excitation band of GLB resulting in a decline in its fluorescence signal (λexc = 234 nm, λem = 350 nm). No changes in the fluorescence lifetime of GLB (0.29 ns) were observed in the presence of CF up to 127.6 mg L−1 concentration. The parameters that affect the fluorescence intensity were investigated, such as fluorophore concentration (16 mgL1), pH (3.2) and temperature (25 °C). Under optimized conditions, the IFE-based approach can determine CF in a range between 1.00 and 100.0 mg L−1, with a detection limit (LOD) of 0.10 mg L−1. The relative standard deviation (% RSD) values for the intra-day and inter-day precision were 0.75 and 1.24, respectively. The new method was successfully tested in the determination of the target analyte in beverage samples without previous treatment. The results were compared with those obtained by a reference method, leading to the conclusion that there were no significant differences at the studied confidence level (α=0.05).