INVESTIGADORES
WUILLOUD Rodolfo German
artículos
Título:
On-line adsorption/preconcentration using a knotted reactor for total Hg determination in drinking water by flow injection CV-AAS
Autor/es:
ALTAMIRANO, JORGELINA C.; WUILLOUD, RODOLFO G.; GÁSQUEZ, JOSÉ A.; OLSINA, ROBERTO A.; MARTINEZ, LUIS D.
Revista:
ATOMIC SPECTROSCOPY
Editorial:
PERKIN ELMER
Referencias:
Año: 2001 vol. 22 p. 398 - 404
ISSN:
0195-5373
Resumen:
An on-line mercury preconcentration and analysis system using atomic absorption spectrometry (AAS) and flow injection (FI) with cold vapor (CV) generation was studied. The mercury was retained as a mercury-2-(5-bromo-2-pyridylazo)-5-diethyla minophenol [Hg(II)-(5-Br-PADAP)] complex, at pH 9.2. The mercury complex was removed from the knotted reactor (KR) with 3 mol/L hydrochloric acid. An enrichment factor of 25 was obtained with respect to CV-AAS. The detection limit for the preconcentration of 25 mL aqueous solution was 5 ng/L. The precision, calculated at peak height for 10 replicate determinations at the 1 ?g/L-Hg level, was 2.8% relative standard deviation (RSD). The calibration curve using the preconcentration system for mercury was linear with a correlation coefficient of 0.9995 at levels near the detection limits up to at least 100 ?g/L. The method was successfully applied to the determination of mercury in drinking water samples.