INQUISUR   21779
INSTITUTO DE QUIMICA DEL SUR
Unidad Ejecutora - UE
artículos
Título:
“In-situ” antimony film electrode for the determination of tetracyclines in Argentinean honey samples
Autor/es:
DI NEZIO, MARÍA S.; KREPPER, GABRIELA; PISTONESI, MARCELO F.; PIERINI, GASTÓN D.
Revista:
SENSORS AND ACTUATORS B-CHEMICAL
Editorial:
ELSEVIER SCIENCE SA
Referencias:
Año: 2017 vol. 241 p. 560 - 566
ISSN:
0925-4005
Resumen:
The main goal of this work is the application of an antimony film electrode (SbFE) prepared for the first time ?in-situ? to determine organic compound as tetracyclines commonly used in Argentine for hive treatment. The SbFE was prepared on a glassy carbon electrode (GCE) and square wave cathodic stripping voltammetry (SWCSV) technique was used. The SWCSV parameters were optimized based on a Draper Lin small composite design. The SWCSV response was linear in the TCs concentration range from 0.40 to 3.00 μM with a limit of detection (LOD) of 0.15 μM. The proposed method is a good and simple alternative to the determination of tetracyclines in Argentinean honeys samples. Recovery experiments were performed using spiked honey samples with standard deviation values from 0.75% and 9.69%.