A Preliminary Study Of An Active Microvalve For Glaucoma
Autor/es:
FABIO A GUARNIERI; FERNANDO L. SASSETTI; LUCIANO GARELLI
Lugar:
Ft Lauderdale
Reunión:
Conferencia; ARVO meeting; 2012
Institución organizadora:
The Association for Research in Vision and Ophthalmology
Resumen:
Purpose: In this study, the characterisation of a preliminary design of an active glaucoma drainage device is presented. Methods: The valve is actuated by conducting polymers and fabricated by MEMS technology. The characterisation process is performed by numerical simulation using the finite element method and a circuit model, and by in vitro hydrodynamic testing. Results: The hydraulic resistance of the designed GDD varies in the range of 13.08-0.36 mmHg min/μl compared with 3.38-0.43 mmHg min/μl for the Ahmed valve. The maximum displacement of the diaphragm in the vertical direction is 18.9 μm, and the strain in the plane is 2%. Conclusions: The proposed preliminary design allows to control the IOP by varying the hydraulic resistance in a greater range than the existing passive valves, and the numerical simulation facilitates the characterisation and the improvement of the design before its construction, reducing time and costs.